The DRE Echo CO2 provides cost-effective SPO2 and CO2 monitoring with accuracy and durability.
The Echo CO2 ensures effective capnography monitoring for intubated and non-intubated patients for continuous long term monitoring. It is tailored for mechanically ventilated and non-intubated patients.
Features:
5.7 inch high resolution display for easy reading
Lightweight, portable design and user-friendly interface for easy operation
Flexible configurations to meet different clinical needs
Standard SpO2
PR measurement
Respironics Loflo sidestream and CAPNOSTAT 5
ETCO2 mainstream measurement
Real-time parameters measurement display with trend table for easy reviewing
Built-in rechargeable Lithium-ion battery for
10 hours continuous working
Powerful storage capacity
Bi-directional communications with central station by wired or wireless network
Nurse call
Physical Specifications
Size: 200.8mm (L) x 241mm (H) x 189 mm (D)
Weight: 3 kg
Display: 5.7 inches
Color TFT resolution: 640X480
Power Supply: 100-240 VAC, 50/60HZ
Pmax=70VA FUSE T 1.6AL
Battery Type: Lithium-ion
Voltage: 14.8 V DC Capacitance: 4,400 mAh
Working period: Color TFT 480min
Rechargeable period: < 360min
Thermal Recorder: (Optional)
Record Width: 48 mm
Paper Speed: 25mm/s
SpO2
Measuring Range: 0 ~ 100 %
Alarm Range: 0 ~ 100 %
Resolution: 1 %
Accuracy
Adult (including Pediatric)
2% (70%~100% SpO2)
Undefined 0~70% SpO2
Neonate
3% (70%~100% SpO2)
Undefined (0~70% SpO2)
Pulse Rate
Measuring and Alarm Range: 30 ~ 300bpm
Resolution: 1bpm
Accuracy: 3bpm
Data update period: 2s
ISO 9919
Respironics CO2
Mainstream
CO2 Measurement Range
0 to 150 mm Hg,0 to 19.7%,
0 to 20 kPa (at 760 mm Hg)
Rise Time
Less than 60 ms Adult/Infant
Reusable or Single Patient Use
Airway Adapter
CO2 Resolution
0.1 mm Hg — 0 to 69 mm Hg
0.25 mm Hg — 70 to 150 mm Hg
CO2 Accuracy
0-40 mm Hg — 2 mmHg
41 – 70 mm Hg — 5% of reading
71 – 100 mm Hg — 8% of reading
101 – 150 mm Hg — 10% of reading
Above 80 bpm — 12% of reading
CO2 Stability
Short Term Drift: Drift over four hours shall not exceed 0.8 mm Hg max.
Long Term Drift: Accuracy specification will will be maintained over a 120-hour period
Respiration Rate Accuracy
1 breath
Calibration
No routine user calibration required. An airway adapter zero is required when changing to a different style of airway adapter
Water Resistance
IPX4-Splash-proof (sensor head only)
Shock Impact
EN60068-2-6 Sinusoidal Vibration
EN60068-2-27 Shock
EN60068-2-64 Random Vibration
Able to withstand repeated 6 foot drops onto tiled floor while operating
Sidestream
Sample Rate: 50 m/L per minute
CO2 Measurement Range:
0 – 150 mm Hg, 0 to 20 kPa
(at 760 mm Hg). Barometric Pressure
supplied by Host
CO2 Resolution:
0.1 mm Hg 0 to 69 mm Hg
0.25 mm Hg 70 to 150 mm Hg
CO2 Accuracy
0 – 40 mm Hg 2 mm Hg
41 – 70 mm Hg 5% of reading
71 – 100 mm Hg 8% of reading
101 – 150 mm Hg 10% of reading
Above 80 bpm 12% of reading
CO2 Stability
Short Term Drift: Drift over four hours shall not exceed 0.8 mm Hg max.
Long Term Drift: Accuracy specification will be maintained over a 120-hour period
Respiratory Rate Range: 2 to 150 bpm
Respiratory Rate Accuracy: 1 breath
Sample Cell/Filter
Proprietary single patient use sample cell and inline filter are integrated with the sample line which eliminates contamination of the internal system
Nasal Sampling Kits for Non-intubated Patients Adult, pediatric and infant nasal CO2 sampling, nasal CO2 sampling and O2 delivery
Adult and pediatric nasal/oral CO2sampling, nasal/oral CO2 sampling and O2 delivery
On-Airway Adapter KITS for Intubated Patients
Adult/Pediatric with and without dehumidification tubing Pediatric/ Infant, low dead space, with and without dehumidification tubing
Taper meets ISO 5356-1
Sample Kit Hours of Use
Nasal Cannula (all styles) — up to 12 hours
On-Airway Adapter Kits without dehumidification tubing — up to 12 hours