DRE Waveline Pro Touch-Screen Anesthesia Monitor
Eight simultaneous waveforms on a single 15-inch display means you won’t miss critical changes during complex cases.
Quality Management
Facility & Processes
Biomedical Technicians
About the DRE Waveline Pro Touch-Screen Anesthesia Monitor
Eight simultaneous waveforms on a single 15-inch display means you won’t miss critical changes during complex cases. The Waveline Pro Touch-Screen Anesthesia Monitor handles everything from basic vitals to advanced hemodynamics without cluttering your workspace with multiple devices.
Motion-tolerant Masimo SET SpO2 monitoring stays accurate even when patients move, while built-in arrhythmia detection catches irregular rhythms automatically. Worth noting: the system stores up to 72 hours of trending data and prints directly from the integrated 3-channel thermal recorder. HL7 compliance means it’ll talk to your existing EMR without custom interfaces.
Two removable batteries plus internal lithium-ion backup keep monitoring continuous during transport. The touch interface responds quickly, and you can configure alarm limits for each case without diving through menu layers.
Key Features
- 15-inch touch-screen displaying 8 waveforms simultaneously
- Motion-tolerant Masimo SET SpO2 with arrhythmia detection
- 3-lead or 5-lead ECG monitoring
- Built-in thermal recorder with 3-channel printing
- 72-hour patient data trending and storage
- HL7-compliant networking via Ethernet and RS-232
- Triple battery system for uninterrupted monitoring
- Optional anesthesia gas bench, EtCO2, and invasive BP modules
- Color-coded alarms with user-configurable limits
Specifications for DRE Waveline Pro Touch-Screen Anesthesia Monitor
Performance
- Display: 15-inch color touch-screen
- Trace: 8 waveforms
- Trend Time: 1 – 72 hour
- Recorder: Built-in thermal array, 3 channels
- Record Width: 48 mm
- Recorder Paper: 50 mm
- Record Speed: 25 mm/s, 50 mm/s
ECG
- Input: 5-lead ECG cable and standard AAMI line for connection
- Lead Choice: I, II, III, aVR, aVF, aVL, V, V1-V6, TEST
- Gain Choice: x0.5, x1, x2, x4
- Frequency Characteristic: 0.05 ~ 35 Hz (+3dB)
- ECG Waveforms: 7 channels
- Sweep Speed: 12.5, 25, 50 and 100 mm/sec
- HR Display Range: 30 ~ 300 bpm
- Accuracy: ±1 bpm or ±1%, whichever is greater
- Alarm Limit Range Setting: Upper 100 ~ 200 bpm, Lower 30 ~ 100 bpm
RESP
- Measure Method: RA-LL impedance
- Range: 0 ~ 120 rpm
- Accuracy: ±3 rpm
- Alarm Limit Setting: Upper 6 ~ 120 rpm, Lower 3 ~ 120 rpm
NIBP
- Measuring Technology: Automatic oscillating measurement
- Cuff Inflating: <30 s (0 ~ 300 mmHg, standard adult cuff)
- Measuring Period (AVE): <40 s
- Mode: Manual, Auto
- Measuring Interval in AUTO Mode: 2 min ~ 4 hrs
- Pulse Rate Range: 30 ~ 250 bpm
- Measuring Range (Adult/Pediatric): SYS 40 ~ 250 mmHg, DIA 15 ~ 200 mmHg
- Measuring Range (Neonate): SYS 40 ~ 135 mmHg, DIA 15 ~ 100 mmHg
- Accuracy: Maximum Mean Error ±5 mmHg; Maximum Standard Deviation 8 mmHg
TEMP
- Range: 25 ~ 50 °C
- Accuracy: ±0.2°C (25.0 ~ 34.9°C); ±0.1°C (35.0 ~ 39.9°C); ±0.2°C (40.0 ~ 44.9°C); ±0.3°C (45.0 ~ 50.0°C)
- Display Resolution: 0.1°C
- Alarm Limit Setting: Upper 0 ~ 50°C, Lower 0 ~ 50°C
- Channel: 2 channels
SpO2 (Masimo SET)
- SpO2 Measuring Range: 1 ~ 100 %
- PR Measuring Range: 25 ~ 240 bpm
- SpO2 Accuracy (non-motion) Adult/Pediatric: 70–100%: ±2%
- SpO2 Accuracy (non-motion) Neonate: 70–100%: ±3%
- SpO2 Accuracy (motion) Adult/Pediatric: 70–100%: ±3%
- Low Perfusion Performance: >0.02% Pulse Amplitude and % Transmission >5
- PR Accuracy (non-motion): 25–240 bpm: ±3 bpm
- PR Accuracy (motion): 25–240 bpm: ±5 bpm
IBP
- Measurement Range: -50 ~ 300 mmHg
- Channel: 2 channels
- Pressure Transducer Sensitivity: 5 µV/V/mmHg
- Impedance Range: 300 ~ 3000 Ω
- Transducer Sites: ART, PA, CVP, RAP, LAP, ICP
- Unit: mmHg/kPa selectable
- Resolution: 1 mmHg
- Accuracy: ±1 mmHg or ±2%, whichever is greater
- Alarm Range: -10 ~ 300 mmHg
EtCO2 (Optional)
- CO2 Measurement Range: 0 ~ 99 mmHg
- Accuracy: ±2 mmHg (0 ~ 38 mmHg); 39–99 mmHg ±5% of reading +0.08% per 1 mmHg above 38 mmHg
- Sampling Rate: 50 ml/min
- Initialization Time: 30 seconds (typical)
- Respiration Rate: 0 ~ 150 breaths/min
- Mode: Adult, Neonate
Cardiac Output (Optional)
- Measurement Method: Thermodilution Method
- C.O. Measurement Range: 0.1 to 20 L/min
- Resolution/Accuracy: C.O. ±5% or ±0.1 L/min, whichever is greater
- Repeatability: C.O. ±2% or 0.1 L/min, whichever is greater
- TB, TI Resolution: 0.1
Anesthetic Agents (Optional)
- Method: Infrared absorption
- Gas Sorts: Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane, CO2, N2O, O2 (optional Automatic Agent ID)
- Measurement Range: Halothane/Isoflurane 0 – 8.5%; Enflurane/Sevoflurane 0 – 10%; Desflurane 0 – 20%; CO2 0 – 10%; N2O/O2 0 – 100%
- Bias/Accuracy: Halothane/Isoflurane/Enflurane/Sevoflurane/Desflurane: ±(0.15 Vol% + 15% rel.); CO2: ±(0.5 Vol% + 12% rel.); N2O: ±(2 Vol% + 8% rel.); O2: ±3 Vol%
Networking
- Wireless: Industry standard 802.11b/g
- Connectivity: Ethernet and RS-232 with open communication protocol
- Compliance: HL7 compliant
Power
- AC Power: 100 – 240 VAC, 50/60 Hz
- Battery: Built-in and rechargeable lithium ion
- Operating Time: 3+ hours
Environmental
- Operating Temperature: 5° – 40° C
- Storage Temperature: -20° – 65° C
- Humidity Operating: ≤ 80%
- Humidity Storage: ≤ 80%
Other
- Other Standard Features: OxyCRG, drug dose calculation, cascading ECG, on-screen NIBP trends (up to 250 readings), user-set defaults, arrhythmia detection, ST segment
- Condition: New
- Category: Patient Monitoring
- Medical Equipment: Anesthesia Monitors