Sapphire Infusion System

Six distinct infusion modes in one compact pump means your staff won’t need to juggle different devices for continuous drips, PCA protocols, TPN.

New List Price $1.00
Your Savings
DRE Price $1.00

Our equipment specialists are here to help you with your questions.

Hours: Monday – Friday 8AM – 5PM ET

Six distinct infusion modes in one compact pump means your staff won’t need to juggle different devices for continuous drips, PCA protocols, TPN administration, and epidural deliveries. The Sapphire handles everything from basic IV fluid delivery to complex multi-step medication protocols with programmable presets that cut setup time.

The system logs up to 8,000 infusion events with multiple authorization levels and confirmation screens built in. Air bubble detection works on both individual bubbles and accumulated air, while the full-color touchscreen handles software updates without additional hardware. Worth noting: the 24-hour battery life means fewer mid-procedure power swaps during long cases.

At under 15 ounces, it’s light enough for mobile applications but stable enough for pole mounting with the snap-on system. Quick priming gets lines ready fast, and the touchscreen interface doesn’t require extensive retraining for staff familiar with modern devices.

Key Features

  • Six infusion modes: continuous, intermittent, TPN, PCA, multi-step, and epidural
  • Full-color touchscreen with programmable presets
  • 24-hour rechargeable battery life
  • Snap-on, snap-off mounting system
  • Air bubble detection for individual and accumulated air
  • 8,000-event infusion history log
  • Multiple authorization levels with confirmation screens
  • Quick priming capability

Physical

  • Dimensions: 5.63” (14.3 cm) H x 3.78” (9.6 cm) W x 1.93” (4.9 cm) D
  • Weight: 14.7 oz (0.4 kg) without battery

Power

  • Battery Life: 24 hours at 125 mL/hr
  • Power Requirements: 100-240V, 50-60Hz, 0.6 A

Delivery

  • Volume Range: 1 – 9999 mL (increments of 1 mL)
  • Delivery Rate Accuracy: ± 2.5% (subject to external conditions)
  • Delivery Rates: 0.1 – 99.9 mL/hr (increments of 0.1 mL/hr); 100 – 999 mL/hr (increments of 1 mL/hr)
  • KVO Rate: Up to 20 mL/hr (increments of 0.1 mL/hr)

Alarms

  • System Alarms: Air in Line; Cassette Misplaced; Occlusions; Check for Occlusion; Downstream Occlusion; Upstream Occlusion; Flow Error; Insufficient Battery
  • Occlusion Alarm Pressure: Up to 17.4 PSI (1.2 Bar)

Materials

  • Housing: Lexan (durable, damage/dust/fluid resistant)

Compatibility

  • Cassette Door: Not integral to pumping process

Intended Use

  • Applications: Controlled infusion delivery in clinical settings
Certified Refurbished