Call Us Today

(502) 244-4444

Certified Refurbished

Valleylab Force EZ Electrosurgical Generator

The Force EZ’s Instant Response™ technology adjusts power output in real time, maintaining consistent cutting and coagulation effects even when tissue.

New List Price $1.00
Your Savings
DRE Price $1.00

Our equipment specialists are here to help you with your questions.

Hours: Monday – Friday 8AM – 5PM ET

The Force EZ’s Instant Response™ technology adjusts power output in real time, maintaining consistent cutting and coagulation effects even when tissue impedance changes mid-procedure. Unlike older generators that deliver fixed power regardless of tissue conditions, this feedback system prevents the power drops and inconsistent results that can occur with varying tissue types.

Three internal microcontrollers process tissue feedback faster than single-processor units, allowing the system to react within milliseconds to maintain surgical effect. The upside is improved performance at lower power settings, which reduces unwanted tissue damage and neuromuscular stimulation during delicate procedures.

The rear panel includes separate monopolar and bipolar footswitch connections, plus a front-mounted switchable connector for quick setup changes. Direct cable integration with Force Argon II systems and OptiMumm smoke evacuators eliminates compatibility guesswork.

Key Features

  • Instant Response™ feedback technology maintains consistent power delivery regardless of tissue impedance
  • Three microcontrollers enable faster system reaction and processing
  • Pure Cut and Blend modes with real-time power adjustment
  • Low and High coagulation modes for contact and noncontact applications
  • Standard bipolar mode uses low voltage to minimize sparking
  • Multiple footswitch connectors including front-mounted switchable option
  • Direct compatibility with Force Argon II and OptiMumm systems

Physical

  • Height: 5 inches (12.7 cm)
  • Width: 16 inches (40.6 cm)
  • Length: 15.6 inches (39.5 cm)
  • Weight: < 15 lbs (6.8 kg)

Output Waveforms

  • Bipolar: 470 kHz sinusoid
  • Monopolar Cut — Pure: 393 kHz sinusoid
  • Monopolar Cut — Blend: 393 kHz bursts of sinusoid recurring at 27 kHz intervals, 50% duty cycle envelope

Monopolar Coagulation

  • Desiccation Low 1: 240 ± 40 kHz sinusoid recurring at 39 kHz, 8% duty cycle
  • Desiccation Low 2: 393 kHz sinusoid
  • Desiccation Low 3: 393 kHz sinusoid
  • Fulguration High 1: 470 ± 40 kHz damped sinusoidal bursts, repetition frequency 57 kHz
  • Fulguration High 2: 470 ± 40 kHz damped sinusoidal bursts, repetition frequency 30 kHz

Output Characteristics

  • Bipolar (Standard Mode) — Maximum P-P Voltage: 300; Rated Load: 100 Ω; Maximum Power: 70 W; Crest Factor (Typical): 1.5
  • Monopolar Cut (Pure) — Maximum P-P Voltage: 2000; Rated Load: 300 Ω; Maximum Power: 300 W; Crest Factor (Typical): 1.5
  • Monopolar Cut (Blend) — Maximum P-P Voltage: 3400; Rated Load: 300 Ω; Maximum Power: 200 W; Crest Factor (Typical): 2.1
  • Monopolar Coag Desiccate Low 1 — Maximum P-P Voltage: 3500; Rated Load: 500 Ω; Maximum Power: 120 W; Crest Factor (Typical): 5.0
  • Monopolar Coag Desiccate Low 2 — Maximum P-P Voltage: 660; Rated Load: 300 Ω; Maximum Power: 120 W; Crest Factor (Typical): 1.5
  • Monopolar Coag Desiccate Low 3 — Maximum P-P Voltage: 1100; Rated Load: 300 Ω; Maximum Power: 120 W; Crest Factor (Typical): 1.5
  • Monopolar Coag Fulgurate High 1 — Maximum P-P Voltage: 6200; Rated Load: 500 Ω; Maximum Power: 120 W; Crest Factor (Typical): 5.0
  • Monopolar Coag Fulgurate High 2 — Maximum P-P Voltage: 8500; Rated Load: 500 Ω; Maximum Power: 120 W; Crest Factor (Typical): 7.0

Performance

  • Power Efficiency Rating (PER): 97
  • Output power variation: Changes by less than 15% or 5 watts (whichever is greater) as line voltage varies from 90–135 V into a 300 Ω load

Leakage and Safety

  • Low Frequency Leakage (50–60 Hz) — Normal polarity, intact chassis ground: < 10 µA; Normal polarity, ground open: < 50 µA; Reverse polarity, ground open: < 50 µA; Sink current, 140 V applied, all inputs: < 50 µA
  • High Frequency Leakage — Bipolar: < 60 mArms; Monopolar: < 150 mArms

REM Contact Quality Monitoring

  • Measurement Frequency: 80 kHz ± 10 kHz
  • Measurement Current: < 10 µA
  • Acceptable Resistance Ranges: REM pad 5–135 Ω; Non-REM pad < 20 Ω
  • Adaptive REM Trip: Baseline impedance plus 40%

Input Power Requirements

  • 110–120 Vac Operating Range: 85–140 Vac; Max Current Cut: 8 A; Max Current Coag: 4.2 A
  • 220–240 Vac Operating Range: 170–280 Vac; Max Current Cut: 4 A; Max Current Coag: 2.1 A

Compatibility

  • Compatible Devices: Force Argon II, Force GSU argon coagulation system, OptiMumm smoke evacuator