Tuttnauer 5050 ELV Fully Automatic Vertical Sterilizer

The Tuttnauer 5050 ELV runs six preset sterilization programs automatically, handling everything from liquid media to glassware without manual intervention.

New List Price $1.00
Your Savings
DRE Price $1.00

Our equipment specialists are here to help you with your questions.

Hours: Monday – Friday 8AM – 5PM ET

The Tuttnauer 5050 ELV runs six preset sterilization programs automatically, handling everything from liquid media to glassware without manual intervention. Its microprocessor monitors pressure and temperature through a PT-100 sensor in a reference vessel, confirming actual sterilization conditions rather than just chamber readings.

The LCD display tracks cycle progress in real-time, showing pressure, temperature, remaining time, and any alerts. Worth noting: the RS-232 interface lets you integrate cycle data into your lab management system, while validation ports accept external temperature sensors for qualification runs.

Multiple safety interlocks prevent door opening under pressure or high temperature. The exhaust steam gets cooled before discharge, and the heat-resistant plastic lid covering prevents accidental contact with hot surfaces.

Key Features

  • Six preset programs with microprocessor control
  • PT-100 temperature sensor in reference vessel for accurate monitoring
  • LCD display with real-time pressure, temperature, and time readouts
  • Automatic demineralized water level regulation
  • RS-232 interface for data integration
  • Validation ports for external temperature sensors
  • Multiple door safety interlocks and pressure relief valve
  • Cooled exhaust steam discharge

Capacity

  • Chamber Volume: 110 L

Sizes

  • Chamber Dimensions: 500 × 500 mm (Ø × depth)
  • External Dimensions: 820 (H) × 880 (W) × 700 (D) mm

Materials

  • Components: Plastic; Stainless Steel

Electrical

  • Heating Performance: 9.0 KW
  • Voltage: 380-400 V, 50/60 Hz

Intended Use

  • Applications: Liquids (nutrient media, buffer solutions), solid items (pipettes, tubes, filters), glassware and plastic articles, waste sterilization