The GE Datex Ohmeda Tec 7 delivers consistent anesthetic agent concentrations across wide flow ranges, maintaining steady output whether you’re running low-flow or high-flow cases. Unlike older vaporizer designs that drift at extreme flows, the Tec 7 holds its concentration settings reliably.
The large dial gives you fine control over concentration adjustments, with clear gradations that make precise titration straightforward during cases. The factory-service free design means fewer maintenance headaches for your biomed team.
Worth noting: the Easy-fil system reduces spillage during refilling, and sevoflurane units get the Quik-fil option for faster turnaround between cases.
Key Features
- Available for isoflurane, sevoflurane, enflurane and halothane
- Easy-fil filling system minimizes leak risk
- Quik-fil option available for sevoflurane models
- Large ergonomic dial with fine gradations
- Factory-service free design
- Consistent output across wide flow ranges
Specifications for GE Datex Ohmeda Tec 7 Anesthetic Vaporizer
Capacity
- Liquid Capacity (full, dry wicks): 300 ml (nominal)
- Retained by wick system: 75 ml (nominal)
- To fill from minimum to maximum mark: 5% Vaporizer: 170 ml (nominal)
- To fill from minimum to maximum mark: 8% Vaporizer (Sevotec): 137 ml (nominal)
Weight
- Empty: 15 lbs 6 oz (7 kg)
Dimensions
- Depth: 8.25 in (210 mm)
- Width: 4.375 in (110 mm)
- Height (unlocked position): 9.875 in (250 mm)
Performance
- Output Range: 200 mL/min to 15 L/min
- Accuracy at 5 L/min O2, 21 ± 2 °C — Vaporizer 5%: ± 0.25% of delivered agent or ± 15% dial setting (whichever is greater)
- Accuracy at 5 L/min O2, 21 ± 2 °C — Vaporizer 8%: ± 0.4% of delivered agent or ± 15% dial setting (whichever is greater)
- Flow Resistance at 5 L/min O2, 21 ± 2 °C: 10 – 15 cm H2O with Vaporizer setting ON +0%
Operating Conditions
- Operating Temperature Range: 18 °C to 35 °C (64 °F to 95 °F)
- Storage Temperature Range: –40 °C to 65 °C (–40 °F to 149 °F)
- Note: Protect the vaporizer packaging from condensation.
Materials
- Components: Plastic
Illumination
- Type: Conventional; Fiber Optic