Drager Air-Shields Isolette C2000 Infant Incubator
The Drager Air-Shields Isolette C2000 uses patented dual air curtains to maintain stable temperatures even when access doors are opened, reducing the.
Quality Management
Facility & Processes
Biomedical Technicians
About the Drager Air-Shields Isolette C2000 Infant Incubator
The Drager Air-Shields Isolette C2000 uses patented dual air curtains to maintain stable temperatures even when access doors are opened, reducing the thermal stress that disrupts fragile neonates during routine care. Unlike standard incubators that lose significant heat with each access, this unit minimizes temperature fluctuations that can compromise patient stability.
Built-in trending tracks air temperature, skin temperature, oxygen levels, humidity, and weight changes over 24-hour periods, giving you the data continuity that’s often lost during shift changes. The optional in-bed scale lets staff monitor weight gain without removing infants from their controlled environment. Worth noting: the large mattress supports twin care when bed space is tight.
Dual access doors mean care teams can work from either side without repositioning equipment or disrupting the infant. Internal noise stays at 47 dBA or lower.
Key Features
- Patented dual air curtains reduce heat loss during door access
- 24-hour trending for temperature, oxygen, humidity, and weight
- Optional dual access doors for workflow flexibility
- Large mattress supports twin care with dual skin probes
- Optional in-bed scale eliminates environment disruption for weighing
- Maintains ≤47 dBA noise levels
Specifications for Drager Air-Shields Isolette C2000 Infant Incubator
Contents
- Standard Hood Includes: 4 port doors; 2 iris ports; front access panel; 3 left and 3 right tubing grommets-front
Sizes
- Mattress Tray Size: 16″ x 32″ (40.6 x 81 cm)
- Mattress to Hood: 16.25″ (41.2 cm)
- Access Panel Opening: 11″ (28 cm)
Capacity
- Scale Weight Range: 0 to 15.4 lbs (0 to 7 kg)
Materials
- Components: Stainless Steel
Illumination
- Illumination: Conventional; Fiber Optic
Compatibility
- Outputs: RS-232
- AC Outlets: 6 AC (Mains) outlet
Weight
- Air-Shields Incubator with VHA Stand Weight: 198 lbs (90 kg)
- Air-Shields Incubator with Fixed Height Stand Weight: 175 lbs (79.5 kg)
Dimensions
- VHA Stand Width: 38.0″ (99 cm); Depth: 26.3″ (67 cm); Height: 52.5″ – 60″ (133.4 to 152.4 cm)
- Fixed Height Stand Width: 46.8″ (119 cm); Depth: 26.3″ (67 cm); Height: 56″ (142 cm)
Performance
- Air Mode Control Range: 20 to 39.0°C
- Skin Mode Control Range: 34.0 to 38.0°C
- Temperature Rise Time (at 22°C ambient): <35 minutes
- Air Flow Velocity Across Mattress: <10 cm/sec
- Micro Air Intake Filter Efficiency: 99.9% for 0.3 micron particles
- CO2 Levels: <0.5% when delivered at 750 ml/min at a point 10 cm above the mattress
Alarms
- Alarms: High/Low Skin Temperature; High/Low Air Temperature; High Temperature Cut-out; Air Probe Failure; Skin Probe Failure/Disconnect; Low Air Flow; Low Humidity (humidity option); Low/High Oxygen (oxygen option); Oxygen Calibration Required (oxygen option); Oxygen Cell Difference (oxygen option); Oxygen Solenoid Failure (oxygen option); Scale Disconnect (scale option); Too Much Weight (scale option); Controller Failure; Sensor Module Failure; Motor Failure; Power Failure; Heater Failure; Sensor Disconnected; Key Stuck
Options
- Servo Humidity Option: Water Capacity: 1000 ml; Refill Time: >24 hrs @ 85% RH; Range: 30 to 95% RH; Display Accuracy: ±6% RH
- Servo Oxygen Option: Accuracy 100% cal ±3%; 21% cal ±5%; Range: 21 to 65%; Resolution: 1%
- Scale Option: Weight Display Resolution: 1 g or 1 oz; Weight Accuracy: 2 g ± 1/2 digit up to 2 kg; 5 g ± 1/2 digit over 2 kg; Weight Gain/Loss Trending Over 7 Days
- Other Options: IV Pole (10.4 lb / 4.72 kg limit); Monitor Shelf (25 lb / 11.4 kg limit); Rear Access Door; Ventilator Tube Support; Oxygen Tank Mount; Drawer Options (large deep swivel; two shallow swivel out)
Controls & Display
- Controller: LCD with brightness control
- Selectable Color Combinations: White on blue (default) or yellow on black
- Keypad Lock: Yes
- Dual-skin Temperature Monitoring: Yes
- 24 Hour Trend Of: Air Temperature; Skin Temperature (1 and 2); Relative Humidity; Oxygen Concentration; Heater Power
Environmental Requirements
- Operating Temperature Range: 20°C to 30°C ambient
- Storage Temperature Range: -25°C to 60°C ambient
- Operating Humidity Range: 5% RH to 99% RH, non‑condensing
- Storage Humidity Range: 0% to 99% RH, non‑condensing
- Operating RH Sensor Temp. Range: 20°C to 41°C
- Operating O2 Sensor Temp. Range: 20°C to 41°C
- CO2 Sensor Calibration Pressure: 600 to 900 mm Hg
Noise
- Operating Noise Level in Hood: 47 dBA (49 dBA with Servo Controlled Oxygen)
Intended Use
- Applications: Provide a heated and humidified environment for neonatal care and thermal support of low birth weight neonates