Call Us Today

(502) 244-4444

New

Penlon Sigma Delta Anesthetic Vaporizer

The Sigma Delta maintains consistent anesthetic agent delivery even when fresh gas flows drop to minimal levels, where many vaporizers start losing accuracy.

New List Price $1.00
Your Savings
DRE Price $1.00

Our equipment specialists are here to help you with your questions.

Hours: Monday – Friday 8AM – 5PM ET

The Sigma Delta maintains consistent anesthetic agent delivery even when fresh gas flows drop to minimal levels, where many vaporizers start losing accuracy. This stability at low flows means you can reduce agent waste while maintaining precise concentration control throughout long procedures.

Worth noting: the vaporizer accommodates four different back-bar mounting systems, so it’ll integrate with your existing anesthesia machines whether you’re running Selectatec, Drager Plug-In, North American Drager, or Cagemount setups. The multiple agent filler blocks let you switch between sevoflurane, isoflurane, halothane, and enflurane without cross-contamination concerns.

At under 11 pounds, it’s lighter than most comparable vaporizers. Both key fill and pour fill models available depending on your workflow preferences.

Key Features

  • Stable performance at low fresh gas flows
  • Compatible with four major back-bar mounting systems
  • Multiple agent filler blocks available (Sevoflurane, Isoflurane, Halothane, Enflurane)
  • Lightweight construction for easy handling
  • Key fill or pour fill options

Weight

  • Weight: 4.8 kg approx.

Sizes

  • Cagemount Dimensions: 133 mm wide × 158 mm deep × 219 mm high
  • Selectatec® Compatible Dimensions: 120 mm wide × 190 mm deep × 242 mm high
  • Drager Plug-In® Compatible Dimensions: 100 mm wide × 190 mm deep × 242 mm high

Capacity

  • Volume at MAX mark: 250 mL nominal
  • Volume at MIN mark: 35 ±10 mL
  • Retained by wick after draining: approximately 60 ±10 mL

Performance

  • Flow Range: Operating flow range 0.2 to 15 litres/min
  • Temperature Range: Operating temperature range 15° to 35°C (58° to 95°F)

Materials

  • Components: plastic; stainless steel

Compatibility

  • Agent Options: Sevoflurane; Isoflurane; Halothane; Enflurane
  • Mounting Options: Selectatec®; Drager Plug-In®; North American Drager; Cagemount

Service

  • Service Interval Guidance: Presumes ten year product life requiring no preventive maintenance service; it is recommended that a service is carried out at ten years. Halothane vaporizers require a five year service.