Tuttnauer Large Laboratory Sterilizer

Tuttnauer Large Laboratory Sterilizer

Brand: Tuttnauer
Type: Autoclaves
Web ID: 889

0 item is in your cart. View cart now

Need help choosing?

Ask an Equipment Expert

(502) 244-4444

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

Details

The ever-growing demand for more efficient and advanced industrial and scientific equipment with higher capacity and better performance, providing safe and effective sterilization, has led to the development of the Tuttnauer's line of Large Sterilizers for Laboratories.

The Large Laboratory Line is designed to cover a wide range of applications for Laboratories, Biotech Facilities and Research Institutes. The Large Laboratory Line operates with saturated steam as the sterilization agent; working at pressures of 0-350 KPa abs and temperatures in the range of 105° C to 137° C.

Special temperature settings are available on request. The sterilization method substantially shortens the sterilization cycle, achieved by evacuating the air with a vacuum pump, combined with steam pulsing. A 0.2µm HEPA Filter is provided to filtrate the air, which equalizes the vessel pressure after the drying cycle.

Tuttnauer offers a large range of sterilizer models, ranging from 120L chamber sizes to 1015L, according to the specific use and nature of product requirements.

Standards

The Large Laboratory Sterilizer Line meets the provisions of the Medical Device Directive 93/42EEC as well as the Pressure Equipment Directive 97/23EEC.

The Horizontal Sterilizer complies with EN285 and EN554 as well as Health Technical Memorandum 2010 and 2031. In addition, the Quality System is certified in compliance with ISO 9002, EN 46002 and ISO 13485. The devices are UL Listed, FDA declared and CE marked (MDD & PED).

Sterilizer Construction

Pressure Vessel and Jacket

The Large Laboratory Line's pressure vessel and jacket is constructed of 316L grade stainless steel.

Doors

The sterilizers' doors are constructed of heavy-duty 304 grade stainless steel. The sterilizer can be supplied with one or two doors. Door options are various: Fully automatic, vertical sliding, horizontal sliding, or hinged with automatic locking.

Door Safety Features

The safety features prevent the operator from opening all door types in the following circumstances:

  • When chamber is pressurized
  • If temperature in the chamber is above safety level
  • For double door models: If one door is open
  • The automatic sliding door's progress will be automatically stopped as a result of any obstruction

Piping

The pipes and fittings are of stainless steel and brass. The valves are ball type, self cleaning and pneumatically operated.

Mounting

The sterilizer can be supplied in one of the following mounting alternatives: Built -in wall mounted or free-standing with cabinet.

Control Monitoring

The sterilizer's Operator Interface Panel is composed of the following:

  • The operation panel contains 2 LEDs. The RUN indicator (green) lights up during process. The FAIL indicator (red) lights up during system failure.
  • Power On/Stand By Keypad.
  • The illuminated LCD display consists of 4 alphanumeric rows with 20 characters in each row.
  • Multifunction keypads are located under the display. The chosen function is displayed above the keypad. These keys can also operate calibration, setup and door status.
  • Up/Down keypads. In normal working mode these are used to browse through the different cycles. Under setup mode they will adjust numeric values.
  • 10 Digit Keypad. These keypads are used for selecting the cycle number or, under setup mode, to adjust numeric values.
  • Stop/Abort Key. Stops operation at any stage of the cycle. This key is active only while the autoclave is in process. This key also deactivates an alarm.
  • Start Keypad. While the autoclave is in the ready position, pressing the START key will activate the cycle.

The Control and monitoring applications can also be performed using a Touch Screen assisted by the ADMC software, which is optionally available.

Cycles Available

The sterilizer is designed to operate up to (10) ten Standard sterilization programs between 105°C and 137°C and (2) two test programs. All programs are selected by using the numeric keypad, and all programs are programmable. The standard programs comply with the European Norm 285.

  • Bowie & Dick Test: Program 11.
  • Vacuum Leak Test: Program 12.

Air removal, sterilization and drying stages can be reprogrammed.

Printer

For clear and concise documentation of cycle process, the control unit is provided with a 24 character per line printer, which is connected to the microprocessor. The printer generates a hard copy printout of the relevant information concerning the cycle operation such as the temperature, pressure, vacuum, sterilization and drying time, number of cycles, etc. In case of an uncompleted cycle, the printout indicates the cycle failure and the cause of the failure. An additional graphic printer is optional. RS485/RS232 PC and Advanced PC Windows-based software for monitoring, logging and control is also available. A circular or trace strip chart temperature and pressure recorder can be supplied on request.